TR 080A – diameter 4 mm, -30 °C to 180 °C

These resistance temperature sensors are designed to measure the temperature of gaseous, liquid or solid substances. The maximum temperature range of use of the sensors is -30 to 180 °C. Used type of lead-in case has silicone insulation and shielding. The sensors are primarily designed for measuring the temperature in air-conditioning ducts. The 4 mm case diameter enables quicker response to changes in temperature. the sensors are designed for use in a chemically non-aggressive environment. The method of use must be chosen with regard to the temperature and chemical resistance of the case and lead-in cable.

Specification

Sensor type TR 080A
Measuring range -30 to 180 °C (can be limited by the sensing element, determine in documentation)
Type of sensing element* Ni 1000, Pt 100, Pt 500, Pt 1000, NTC
Ingress protection IP 65 in accordance with EN 60529, as amended
Thread / OK according to customer
Case material stainless steel DIN 1.4301
Diameter of case 4 mm
Length of case L 30 to 200 mm
Lead-in cable shielded silicone 2 x 0.22 mm2
shielded silicone 4 x 0.15 mm2
Wire resistance 0.16Ω for 1 m of cable for 2-wire connection
Time response τ0.5 < 5 s (in flowing water 0.4 m.s-1)

*) Certain technical specifications of thermocouple sensors (lead wires, IP rating, etc.) may differ with different types. Note: Manufacturer reserves the right to make changes in design and technical characteristics of the products. For actual information, please always contact your dealer.

Dimensional Draft

Wiring Diagram

Visit our downloads page to view technical characteristics or contact our technical sales department for more information, prices or with your specifications.

Also, you can reach us on +44 (0) 1865400514