These resistance temperature sensors are designed to measure the temperature of gaseous, liquid or solid substances. The maximum temperature range of use of the sensors is -30 to 180 °C. Used type of lead-in case has silicone insulation and shielding. The sensors are primarily designed for measuring the temperature in air-conditioning ducts. The 4 mm case diameter enables quicker response to changes in temperature. the sensors are designed for use in a chemically non-aggressive environment. The method of use must be chosen with regard to the temperature and chemical resistance of the case and lead-in cable.
TR 080A – diameter 4 mm, -30 °C to 180 °C
- Home
- Products
- Sensors and Probes
- Temperature Sensors
- Temperature Sensors with Cable
- Temperature Sensors with a Thread
- TR 080A – diameter 4 mm, -30 °C to 180 °C
Specification
Sensor type | TR 080A |
---|---|
Measuring range | -30 to 180 °C (can be limited by the sensing element, determine in documentation) |
Type of sensing element* | Ni 1000, Pt 100, Pt 500, Pt 1000, NTC |
Ingress protection | IP 65 in accordance with EN 60529, as amended |
Thread / OK | according to customer |
Case material | stainless steel DIN 1.4301 |
Diameter of case | 4 mm |
Length of case L | 30 to 200 mm |
Lead-in cable |
shielded silicone 2 x 0.22 mm2 shielded silicone 4 x 0.15 mm2 |
Wire resistance | 0.16Ω for 1 m of cable for 2-wire connection |
Time response | τ0.5 < 5 s (in flowing water 0.4 m.s-1) |
*) Certain technical specifications of thermocouple sensors (lead wires, IP rating, etc.) may differ with different types. Note: Manufacturer reserves the right to make changes in design and technical characteristics of the products. For actual information, please always contact your dealer.
Dimensional Draft
Wiring Diagram
Visit our downloads page to view technical characteristics or contact our technical sales department for more information, prices or with your specifications.
Also, you can reach us on +44 (0) 1865400514
Also, you can reach us on +44 (0) 1865400514