Home Products STHC 102 – TEMPERATURE, RELATIVE HUMIDITY AND CO2 TO THE INTERIOR WITH RS 485 OUTPUT (MODBUS)
Back

STHC 102 – TEMPERATURE, RELATIVE HUMIDITY AND CO2 TO THE INTERIOR WITH RS 485 OUTPUT (MODBUS)

The STHC 102 temperature, relative humidity and CO2 sensor is designed to measure the carbon dioxide concentration, temperature and relative humidity of the air in spaces protected against water.

The STHC 102 temperature, relative humidity and CO2 sensor consists of a plastic ribbing head where a printed circuit board with the individual sensors and a converter is placed to establish a communication via the RS 485 bus. The temperature and relative humidity are measured by a common internal sensor whose signal is processed in a microprocessor and is converted to an output signal of MODBUS RTU. The CO2 value is measured by a NDIR module whose digital signal is also converted to an output signal of MODBUS RTU. For the CO2 concentration sensor, there is an autocalibration function available to set the sensor at the minimum CO2 value corresponding to the outside concentration level. The STHC 102 temperature, relative humidity and CO2 sensor meets the ingress protection of IP 30 according to EN 60529, as amended. Suitable design and high-quality material ensure that the sensor does not feel disturbing even in the interiors with high aesthetic requirements. You can choose the color of the box white or gray.

Enquire for pricing...
SHARE VIA
Sensor type STHC 102
Temperature measurement range **
0 to 45 °C with guaranteed accuracy of CO2 measurement
– 30 to 70 °C without guaranteed accuracy of CO2 measurement
Temperature measurement accuracy * ± 0,5 °C
Relative humidity measurement range *
0 to 85 % with guaranteed accuracy of CO2 measurement
0 to 95 % without guaranteed accuracy of CO2 measurement
Relative humidity measurement accuracy *
± 3 % in range 10 to 85 %
± 4.5 % in range 0 až 10 %
CO2 measuring range * 400 to 5000 ppm
CO2 measuring accuracy * ± 100 ppm
Time response CO2 (90%) 90 s
Output signal RS 485 / MODBUS RTU
Supply voltage U 15 to 30 VDC
Rated supply voltage Un 24 VDC
Consumption
maximum: 500 mW
typical: 300 mW
Ingress protection IP 30 acc. to EN 60529, as amended
Dimension of the head 71.9 x 59 x 27 mm
Material of the head LEXAN
Weight min 35 g
Recommended wire cross section 0.14 to 1 mm2

The STHC 102 temperature, relative humidity and CO2 sensor is designed to be operated in a chemically non-aggressive environment; its use must be chosen with regard to temperature and chemical resistance of the head and of the individual sensors.

The operating conditions to establish the correct function are:

Ambient temperature in the vicinity of the sensor: 0 to 45 °C
Relative ambient humidity: 0 to 95% (non-condensing humidity)
Atmospheric pressure: 87 to 106 kPa

Customer help

Can’t see what your looking for? Please get in touch if you’d like some direct assistance.

Get in touch